WARNINGS AND SAFETY INSTRUCTIONS

If thermal disinfection is required
(93 °C / 10 min.), the process
needs to be checked and documented
regularly. This can be done by check-
ing the disinfection control diode on the
machine. Alternatively, a chemical indi-
cator may be used (e.g. the Disinfec-
tion Control Tubes by A. Browen Ltd.,
supplied by Miele). These disinfection
control tubes should be placed in the
filter combination. After the disinfection
cycle, they may be inspected by re-
moving the coarse filter.
CAUTION: Washer/disinfector must
not be used without all filters in
place.
CAUTION: Use only accessories
specifically designed for this
equipment. Load accessories into
washer/disinfector in accordance with
the instructions provided.
CAUTION: The operator is respon-
sible for monitoring and guarantee-
ing the standard of disinfection in the
routine disinfection cycle. Disinfection
parameters should be checked regu-
larly using chemical indicators, and
contamination levels checked and doc-
umented periodically using biological
indicators.
SAVE THESE
INSTRUCTIONS