8. Error messages / troubleshooting
In case of faults, the E_ message appears in the display.
Error messages can occur when
1. inflation pressure is higher than 300 mmHg, E2,
2. blood pressure is unusually high or low E3,
3. you move or talk during the measurement (in addition to E_,
the cardiac arrhythmia icon "
4. the cuff tube is not inserted properly E1,
5. the cuff is inflated for longer than 25 seconds E1.
In the above cases, you must repeat the measurement. Make
sure that the cuff tube is properly inserted and that you do
not move or talk. Re-insert the batteries if necessary, or else
replace them.

9. Specifications

Model No.
Method of Measurement Oscillometric, non-invasive blood
Range of Measurement
Accuracy of the
pressure reading
Measuring uncertainty
Sets of Memory
" appears in the display E3),
SBM 30
pressure measurement on the upper
arm
systolic 40 – 280 mmHg,
diastolic 40 – 280 mmHg,
pulse 40 –199 beats/minute
systolic ± 3 mmHg /
diastolic ± 3 mmHg /
pulse ± 5 % of the displayed value
max. permissible standard deviation
according to clinical testing:
systolic: 8 mmHg /diastolic: 8 mmHg
2 x 60 memory places
Unit Dimension
Unit Weight
Cuff Size
Operating Temperature
Storage-and
transportemperature
Power Supply
Battery Life
Accessories
Classification
Protection class
Key to symbols
These specification are subject to change without notice for
purpose of improvement.
• This unit is in line with European Standard EN60601-1-2 and
is subject to particular precautions with regard to electro-
magnetic compatibility (EMC). Please note that portable and
mobile HF communication systems may interfere with this in-
strument. For more details, please contact customer service
at the address indicated.
17
L135 x W105 x H53 mm
Approx. 170 gr.
Cuff for upper arm circumference of
23 to 33 cm
+ 10 °C to + 40 °C, ≤ 85 % RH
- 20 °C to + 70 °C, ≤ 85 % RH
4 x 1,5 V AAA batteries (alkaline type
LR03)
For approx. 250 measurements,
depending on blood pressure level
or inflation pressure
With storage pouch, instruction
manu al, 4 pcs AAA battery
Application part Type BF
internal power supply, IPXO,
no AP or APG, continuous operation
Application part Type BF
Important! Read the operating
instruction!