Saving, Retrieving And Deleting Results; Error Messages/Troubleshooting - Sanitas SBC 23 Instructions For Use Manual

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 13
Blood pressure value
Systole
category
(in mmHg)
High normal
130 – 139 85 – 89
Normal
120 – 129 80 – 84
Optimal
< 120
Source: WHO, 1999 (World Health Organization)

6. Saving, retrieving and deleting results

• The results of each successful measurement are stored together
with date and time. With more than 60 items of measured data,
the earliest items of data measured are lost.
• To recall measurements, press the memory button M. The device
first shows the number of used memory spaces and then changes
automatically to the average of all stored measurements .
• By pressing the memory button M again, the most recent individ-
ual measurements are displayed with the date, time and graphic
WHO classification, while the blood pressure values and the pulse
are displayed in turn.
• In the display of the individual measured values, you have the
option of clearing the memory by pressing the memory button M
for 3 seconds. All values are deleted after 3 short beeps.
Diastole
Action
(in mmHg)
regular monitoring
by doctor
self-monitoring
< 80
self-monitoring

7. Error messages/troubleshooting

In case of faults, the
_ message appears in the display.
Error messages may appear if:
• systolic or diastolic pressure could not be measured (
appears on the display)
• systolic or diastolic pressure was outside the measurement range
( or Lo appears on the display)
• the cuff is fastened too tightly or loosely ( Er3 or Er4 appears on
the display)
• the pump pressure is higher than 300 mmHg ( Er5 appears on the
display)
• pumping up takes longer than 160 seconds ( Er6 appears on the
display)
• there is a system or device error ( ErA ,
on the display)
• the batteries are almost empty
In the above cases, you must repeat the measurement.
Reinsert the batteries if necessary, or else replace them.
Technical alarm – description
Should the recorded blood pressure (systolic or diastolic) lie out-
side the limits specified in the section "Technical specifications", the
technical alarm will appear on the display indicating either " " or
20
or
, Er7 or Er8 appears
.