When the reference trigger occurs, the NI WLS/NET-9163 Series device
continues to write samples to the buffer until the buffer contains the number
of posttrigger samples desired. Figure 10 shows the final buffer.
Reference Trigger
Pretrigger Samples
Posttrigger Samples
Complete Buffer
Figure 10. Reference Trigger Final Buffer
Using a Digital Source
To use ai/ReferenceTrigger with a digital source, specify a source and an
edge. PFI 0 can provide the source. Refer to the Device Routing in MAX
topic in the NI-DAQmx Help or the LabVIEW Help in version 8.0 or later
for more information.
The NI-DAQmx Help is available after installation from Start»Programs»
National Instruments»NI-DAQ»NI-DAQmx Help. To view the
LabVIEW Help, in version 8.0 or later, select Help»Search the LabVIEW
Help in LabVIEW. Alternately, to download the LabVIEW Help, go to
.
ni.com/manuals
You also can specify whether the measurement acquisition stops on the
rising edge or falling edge of ai/ReferenceTrigger.

AI Pause Trigger Signal

The Pause Trigger signal can be generated from external sources. Any time
the signal deasserts, you can use the Pause Trigger signal to pause the
acquisition.
You can use the AI Pause Trigger (ai/PauseTrigger) signal to pause and
resume a measurement acquisition. The internal sample clock pauses while
the external trigger signal is active and resumes when the signal is inactive.
You can program the active level of the pause trigger to be high or low.
Using a Digital Source
To use ai/PauseTrigger, specify a source and a polarity on PFI 0.
NI WLS/ENET-9163 Carrier User Guide and Specifications
20
ni.com