Olympus BX3-SSU Instructions Manual

Scanning stage with ultrasonic
Hide thumbs Also See for BX3-SSU:

Advertisement

INSTRUCTIONS
BX3-SSU
SCANNING STAGE WITH ULTRASONIC
This instruction manual is for the Olympus Scanning Stage with Ultrasonic Model BX3-SSU.
To ensure the safety, obtain optimum performance and to familiarize yourself fully with the use of
the scanning stage with ultrasonic, we recommend that you study this manual thoroughly before
operating the microscope. In addition, be also sure to read the instruction manual for the BX63
microscope on which the scanning stage with ultrasonic is installed.
Retain this instruction manual in an easily accessible place near the work desk for future reference.
A X 7 9 3 4

Advertisement

Table of Contents
loading

Summary of Contents for Olympus BX3-SSU

  • Page 1 BX3-SSU SCANNING STAGE WITH ULTRASONIC This instruction manual is for the Olympus Scanning Stage with Ultrasonic Model BX3-SSU. To ensure the safety, obtain optimum performance and to familiarize yourself fully with the use of the scanning stage with ultrasonic, we recommend that you study this manual thoroughly before operating the microscope.
  • Page 2 Refer to your local Olympus distributor in EU for return and/or collection systems available in your country. NOTE: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules.
  • Page 3: Table Of Contents

    BX3-SSU CONTENTS IMPORTANT -- Be sure to read this section for safe use of the equipment. -- MODULE NOMENCLATURE CONTROLS OPERATION OF SCANNING STAGE WITH ULTRASONIC Turning the Power ON Moving the Stage OPTIONAL ACCESSORIES Stage Center Plate IX-CP50/IX2-GCP Stage Clips IX-SCL...
  • Page 4: Important

    2. To prevent obstruction of the air vents 2 on both sides of the control box @ for BX3-SSU, always install it longitudinally and make sure to leave at least 10 cm of free spaces around them.
  • Page 5: Maintenance And Storage

    (1) Near an air intake or outlet of an air conditioning facility, etc. (2) Place subject to important vibrations or changes in temperature. (3) Near a source (including equipment not manufactured by Olympus) of abnormal noise. (4) Place subject to direct sunlight.
  • Page 6: Module Nomenclature

    MODULE NOMENCLATURE } To ensure the performance, we recommend that you have your Olympus representative assemble and adjust the scanning stage with ultrasonic. The modules marked * are not part of the scanning stage with ultrasonic product. The control of the scanning stage with ultrasonic is possible from the U-MCZ controller, the touch-panel controller provided with the BX63.
  • Page 7: Controls

    For a slide glass sample. Optional accessories Stage clips IX-SCL Center plate with Glass center plate 50 mm hole IX2-GCP IX-CP50 Supersonic stage control box (Provided with BX3-SSU) DRIVE connector RS232C connector SENSOR connector MCXY connector Pilot lamp Main switch TRIG-OUT connector Fuse box (P.
  • Page 8: Controls

    If the foreign substance is entered in the stage driving part, CAUTION the status may not be recovered by performing the Auto Compensation. In this case, turn the power of BX3-SSU Control Box to “ ” (OFF), and perform the Auto Compensation again after moving the stage from front to back and from side to side several times with a hand.
  • Page 9: Operation Of Scanning Stage With Ultrasonic

    BX3-SSU OPERATION OF SCANNING STAGE WITH ULTRASONIC When you use BX3-SSU for the first time, be sure to perform the Auto Compensation of the stage. For detail procedures, refer to “Auto Compensation of the Stage” described in the previous page.
  • Page 10: Optional Accessories

    OPTIONAL ACCESSORIES Stage Center Plate IX-CP50/IX2-GCP (Fig. 5) } When using a dish or petri dish sample instead of a slide glass specimen, remove the slide holder and replace it with the stage center plate. Be careful not to drop or apply a strong impact to the stage CAUTION center plate.
  • Page 11: Specifications

    BX3-SSU SPECIFICATIONS Item Specification Scanning stage with ultrasonic Applicable microscope frame BX63F Slide glass specimen x 2 (total thickness 0.9 to 1.2 mm, cover glass thickness Slide holder 0.17 mm) IX2-SCL. Stage clips (To be attached on two front holes. The rear holes should not be used to avoid interference with the objectives.)
  • Page 12: Troubleshooting Guide

    Under certain conditions, performance of the scanning stage with ultrasonic may be adversely affected by factors other than defects. If problems occur, please review the following list and take remedial action as needed. If you cannot solve the problem after checking the entire list, please contact your local Olympus representative for assistance. Problem...
  • Page 13: Assembly

    · Connect only the designated modules. Connecting a non-designated module may cause malfunction. · Be sure to set the main switches of the BX3-CBH and control box for BX3-SSU to “ ” (OFF) before connecting the cables.
  • Page 14 (Fig. 6) 1. Place BX3-SSU over BX63 so that the respective screw numbers (see the Fig. 6) of BX3-SSU and BX63 are overlapped. 2. Tighten the screw hole @ with the clamping screw (attached with BX3-SSU) from the upper side of BX3-SSU. Use Allen wrench (40x70mm) attached with BX63 for tightening.
  • Page 15 · Cables and cords are vulnerable when bent or twisted. Never subject them to excessive force. · Always use the power cord provided by Olympus. If no power cord is provided, please select the proper power cord by referring to the section “PROPER SELECTION OF THE POWER SUPPLY CORD”...
  • Page 16: Proper Selection Of The Power Supply Cord

    If no power supply cord is provided, please select the proper power supply cord for the equipment by referring to “ Specifications ” and “ Certified Cord ” below: CAUTION: In case you use a non-approved power supply cord for Olympus products, Olympus can no longer warrant the electrical safety of the equipment.
  • Page 17 BX3-SSU Table 2 HAR Flexible Cord APPROVAL ORGANIZATIONS AND CORDAGE HARMONIZATION MARKING METHODS Alternative Marking Utilizing Printed or Embossed Harmoni- Black-Red-Yellow Thread (Length zation Marking (May be located of color section in mm) Approval Organization on jacket or insulation of internal...
  • Page 18 MEMO...
  • Page 20 Printed in Japan on October 12, 2010 M 001-05...