Section 12-Heat Exchanger Assembly; Replacing Heat Exchanger Assembly - Invacare Perfecto 2 SeriesIRC5PO2AW Service Manual

Homefill compatible
Table of Contents

Advertisement

SECTION 12—HEAT EXCHANGER ASSEMBLY
SECTION 12—HEAT EXCHANGER
ASSEMBLY

Replacing Heat Exchanger Assembly

To prevent electrical shock, ALWAYS disconnect concentrator from electrical outlet
before servicing.
NOTE: For this procedure, refer to FIGURE 12.1 on page 59.
NOTE: The heat exchanger assembly contains three components: heat exchanger, nine inch vinyl
tube and connecting component.
1. Unplug unit.
2. Remove cabinet. Refer to Removing Cabinet on page 23.
Use care not to deform heat exchanger when installing, removing or cleaning.
3. Disconnect brass nut that secures the heat exchanger assembly to the compressor
(Detail "A").
4. On the backside of the concentrator, loosen the clamp that secures the vinyl tube to the
manifold assembly and remove the vinyl tube from the manifold assembly. (Detail
"B").
5. Remove heat exchanger assembly by pulling vinyl tube through sound box.
6. Reverse STEPS 3-5 to install new heat exchanger.
7. Run unit and check for leaks. Refer to Leak Test on page 82.
8. Reinstall cabinet. Refer to Removing Cabinet on page 23.
Perfecto
2
DANGER
CAUTION
58
Part No 1154245

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Irc5po2vaw

Table of Contents