TRIO NEBULIZER HANDSET CLEANING
WARNING
To reduce the risk of infection, contamination and/or improper use,
clean the Trio Nebulizer Handset immediately after every use. Do not
wash the Controller, Nebulizer Connection Cord, AC Power Supply or
Plug Converters. Use liquid soap designed for washing dishes. Do not
use "white" liquid dish soaps (e.g. Ivory or Dove) or antibacterial liquid
dish soaps as these may contain additives harmful to the Trio Nebulizer
Handset.
1. Disconnect the Trio Nebulizer Handset from the Nebulizer Connection
Cord (Figure 19).
2. Disassemble the Trio Nebulizer Handset parts:
a. Remove the Medication Cap from the Medication Reservoir
(Figure 20).
b. Remove the Mouthpiece from the Aerosol Chamber (Figure 21).
c. Gently pull up on the tab of the Medication Reservoir. This will
release the Aerosol Chamber so that you may open it (Figure 22).
19
Figure
a
20
Figure
a
21
Figure
22
Figure
b
b
a
TAB
b
d. Remove the Inspiratory Valve from the Aerosol Chamber (Figure 23).
e. Being careful to touch ONLY the plastic outer ring of the Aerosol
Head, press the two flexible plastic tabs on the side of the Aerosol
Head and remove the Aerosol Head (Figures 24-25).
NOTE: There is no need to disassemble the Medication Reservoir from
the Aerosol Chamber either for Cleaning or Disinfecting. Doing so may
damage the hinge that holds them together.
23
Figure
24
Figure
25
Figure
8
|
QUESTIONS ABOUT TRIO? 1-888-335-6946