Pari Trio Instructions For Use Manual page 10

Table of Contents

Advertisement

H
TRIO NEBULIZER HANDSET CLEANING (CONT)
3. To wash the Trio Nebulizer Handset, start by thoroughly rinsing
each of the disassembled parts under warm running tap water for
approximately 10 seconds each.
Rinse the Aerosol Head for at least 10 full seconds on each side (Figure
26). Never use a brush or any other object to clean the AEROSOL HEAD
(Figure 27). THOROUGH RINSING OF BOTH SIDES OF THE AEROSOL
HEAD HELPS TO PREVENT CLOGGING AND ENSURE PROPER OPERATION.
4. Add a few drops of clear liquid dish soap to a clean bowl or basin and
add warm tap water (Figure 28).
5. Wash all parts of the Trio Nebulizer Handset and rinse thoroughly under
running, warm tap water (Figure 29).
If Disinfecting after Cleaning, skip to section I and follow the steps.
26
Figure
27
Figure
28
Figure
QUESTIONS ABOUT TRIO? 1-888-335-6946
6. Air dry the Trio Nebulizer Handset parts on a clean, lint-free towel
(Figure 30).
7. After the Trio Nebulizer Handset parts are completely dry place them
in a bag for storage. Re-assemble the Trio Nebulizer Handset just prior
to the next treatment.
8. If the Controller needs cleaning, wipe down with a slightly damp, soft
cloth (Figure 31). DO NOT allow liquid to get inside the Controller.
29
Figure
30
Figure
31
Figure
|
9
CONTROLLER

Advertisement

Table of Contents
loading

Table of Contents