Technical Specifications
Model-specific Data
Quintix
Model
612-1x
Weighing capacity
g
610
Readability
mg
10
Repeatability
mg
10
(standard deviation)
Linearity
mg
30
Sensitivity drift
± ppm/K 4
between
+10 to +30°C
Typical stabilization
s
1.5
time
isoCAL Settings:
– temperature change
K
2
– time interval
h
6
Display result
s
0.1/0.2
(depending
on the filter level)
Weighing pan size
mm
d 180
Net weight, approx.
kg
4.7
Practum
Model
224-
1x
Weighing capacity
g
220
Readability
mg
0.1
Repeatability
mg
0.1
(standard deviation)
Linearity
mg
0.2
Sensitivity drift
± ppm/K 2
between
+10 to +30°C
Typical stabilization
s
2
time
d 90
Weighing
mm
pan size
Weighing chamber
mm
209
height*
Net weight, approx.
kg
4.5
FCC Rules
Safety Precautions:
NOTE: This equipment has been tested and
found to comply with the limits for a Class B
digital device, pursuant to part 15 of the FCC
Rules. These limits are designed to provide
reasonable protection against harmful
interference in a residential installation. This
equipment generates, uses and can radiate radio
frequency energy and, if not installed and used
in accordance with the instructions, may cause
harmful interference to radio communications.
However, there is no guarantee that interference
will not occur in a particular installation. If this
equipment does cause harmful interference to
radio or television reception, which can be
®
1
1
)
412-1x
)
6101-1x
410
6,100
10
100
10
100
30
300
4
8
1.5
1.5
2
2
6
6
0.1/0.2
0.1/0.2
d 180
d 180
4.7
4.7
®
124-
64-
513-
313-
1
1
1
1
1
)
1x
)
1x
)
1x
)
1x
)
120
60
510
310
0.1
0.1
1
1
0.1
0.1
1
1
0.2
0.2
2
2
2
2
3
3
2
2
1
1
d 90
d 90
d 120 d 120 d 120 d 180 d 180 d 180 d 180 d 180 d 180 d 180 d 180 d 180 d 180
209
209
209
209
4.5
4.5
4.9
4.9
determined by turning the equipment off and on,
the user is encouraged to try to correct the
interference by one or more of the following
measures:
– Reorient or relocate the receiving antenna.
– Increase the separation between the
equipment and receiver.
– Connect the equipment into an outlet on a
circuit different from that to which the
receiver
– Consult the dealer or an experienced radio/TV
technician for help.
Shielded Cables: Connections between the
devices and peripherals must be made using
shielded cables in order to maintain compliance
with FCC radio frequency emission limits.
1
1
1
)
5101-1x
)
2101-1x
)
5,100
2,100
100
100
100
100
300
300
8
8
1.5
1.5
2
2
6
6
0.1/0.2
0.1/0.2
d 180
d 180
4.7
4.7
213-
3102-
2102-
1102-
612-
1
1
1
1
1x
)
1x
)
1x
)
1x
)
1x
210
3,100 2,100 1,100 610
1
10
10
10
10
1
10
10
10
10
2
20
30
30
30
3
4
4
4
4
1
1.5
1.5
1.5
1.5
209
–
–
–
–
4.9
3.1
3.1
3.1
3.1
Installation and Operating Instructions Secura, Quintix, Practum
1
1
6100-1x
)
5100-1x
)
6,100
5,100
1,000
1,000
500
500
1000
1000
8
8
1
1
2
2
6
6
0.1/0.2
0.1/0.2
d 180
d 180
4.7
4.7
412-
6101-
5101-
2101-
1
1
1
1
1
)
1x
)
1x
)
1x
)
1x
)
410
6,100 5,100 2,100 6,100 5,100
10
100
100
100
10
100
100
100
30
300
300
300
4
8
8
8
1.5
1.5
1.5
1.5
–
–
–
–
3.1
3.1
3.1
3.1
Modifications: Any modifications made to
this devices that are not approved by Sartorius
Lab Instruments GmbH & Co. KG may void the
authority granted to the user by the FCC to
operate this equipment.
1
) Possible terms for country-specific models:
x = S: Standard balances without country specific
additions
x = SAR: Standard balances with country specific
additions for Argentina only
x = SJP: Standard balances with country specific
additions for Japan only
x = SKR: Standard balances with country specific
additions for South Korea only
* Upper edge of the weighing pan to the lower
edge of the upper draft shield panel
6100-
5100-
1
1
1x
)
1x
)
1,000 1,000
500
500
1,000 1,000
8
8
1
1
–
–
3.1
3.1
81