Remote Control Instructions - KERN AEJ-C Operating Instructions Manual

Analytical precision balance
Hide thumbs Also See for AEJ-C:
Table of Contents

Advertisement

17.4.2 Remote control instructions

Feedback by weighing scale after sending remote control command:
XX_A
CR LF
XX_I
CR LF
XX_^
CR LF
XX_v
CR LF
XX_E
CR LF
XX
1 – 3
4
Instruction:
1-3
Stabilisation
[blank space]
symbol
[?]
[^]
[v]
[blank space]
Character
9 digits
Weight
3 digits
Unit
AEJ-C/AES-C/PLJ-C-BA-e-1514
Instruction accepted; will be executed
Instruction received; impossible to carry out
Command accepted but max range exceeded
Command accepted but min range exceeded
Error during execution, timeout for stable weighing value
exceeded
Name of command
5
6
7
8 – 16
Character
If stable
If stable
If error (Command + exceeded)
If error (Command – exceeded)
For positive and negative values
With right justifications
With left justification
17
18 - 20
21
22
95

Advertisement

Table of Contents
loading

This manual is also suitable for:

Plj-cAes-c

Table of Contents