Percent Determination - KERN AEJ-C Operating Instructions Manual

Analytical precision balance
Hide thumbs Also See for AEJ-C:
Table of Contents

Advertisement

13.4 Percent determination

Percent determination allows weight display in percent, in relation to a reference
weight. The reference weight may be determined by weighing or by entering it
numerically.
• Determining reference by weighing
The function DETERMINE REFERENCE WEIGHT is allocated to the F1-key.
Go to menu and select P2.6 DEVIATION
Use the arrow keys to select P2.6.6 SHORTCUTS
Press F2, P2.6.6.1 will be displayed along with the previous function for the
F1-key.
Again press F2 and the information line will start flashing.
Use the arrow keys to select DETERMINE REFERENCE WEIGHT and
Confirm by PRINT; this function has now been allocated to F1.
In weighing mode:
(example)
Remove reference weight and place load. The weight will be displayed in
percent according to reference weight.
AEJ-C/AES-C/PLJ-C-BA-e-1514
This completes setting the operating
mode weighing by percentage.
The percent symbol will be
displayed at the upper margin of the
display.
The setting SAMPLE 0.0 G will be
displayed at the lower margin.
Put the reference weight equalling
100% on the weighing plate.
Example: 1 kg is supposed to be
equivalent to 100%.
The reference weight of 1 kg has now been saved
as 100% value.