Rel Test/Touch Hold Test

1. Turn on the GMM to VDC, and apply 1000V dc.
2. Press t [Rel].
3. Check for a Rel reading between the limits shown in Table 4-19.
Input
1000.0V dc
4. Press é.
5. On the 5700A, press l to select Standby.
6. Check for a GMM reading within the limits shown in Table 4-20.
Input
0V dc
4-20
Table 4-19. Rel Performance Test
Input Source
5700A using GMM volts input
Table 4-20. Touch Hold Performance Test
Input Source
(open)
Lower Limit
-0.2
Lower Limit
-0.2
4-17.
Upper Limit
0.2
Upper Limit
0.2