Reasssembly

Generally, reassembly reverses the disassembly procedure. Specifically, do the
following:
1. Verify that the selector switch still points to "OFF".
2. Place the case top face down on a non-marring surface.
3. Grasp the board at the top and bottom so that it can be smoothly lowered into
position in the case top.
4. As you begin lowering the board into position, make sure the battery eliminator
connection (detail "B" in Figure 3-1) slips into its slot snug against the top case
wall.
Gently press the board into position at the top and bottom.
5. Replace the three board-securing screws.
6. Replace the case bottom. Torque the six securing screws to 11-13 inch-pounds in the
sequence shown in Figure 3-2.
At all times, avoid stressing the ribbon cable.
Note
3
Maintenance

Reasssembly