Peak Hold Test; Component Test (867B Only) - Fluke 867B Service Manual

Hide thumbs Also See for 867B:
Table of Contents

Advertisement

Peak Hold Test

Rotate the GMM selector to AC Volts (
Configuration 6 (Figure 4-3). Press u to select Peak Hold. Then make the test called
for in Table 4-17. Peak maximum and minimum readings should be within the specified
limits.
Range
Input
3 VAC
2.828V peak (2.0V rms),
10.0 kHz

Component Test (867B Only)

Use the Fluke 85 in the tests detailed in Table 4-18 to measure the expected limits at the
GMM input jacks (V, e, G, and â). Rotate the GMM selector to Component Test
and measure the appropriate signal. See Figure 4-10.
Range
GMM Output
na
2.2V ac rms, 200 Hz
na
0.210 mA ac rms, 200 Hz
1
2
3
4
FREEZE
HZ
DISPLAY
SAVE
MIN
MODE
PRINT
MAX
WAKE UP
COMPONENT
TEST
A
mA/ A
EXT TRIG
COM
L
Table 4-17. Peak Hold Performance Test
Input Source
5700A
Table 4-18. Component Test Performance Test
Lower Limit
5
RANGE
TOUCH
HOLD
V
Figure 4-10. Configuration 8 (Component Test)
Perfomance Testing and Calibration
). Connect the GMM and the 5700A in
Lower Limit
2.39
1.9
0.164
Performance Tests
4-15.
Upper Limit
3.27
4-16.
Upper Limit
2.5
0.256
85
TRUE RMS MULTIMETER
AUTO
100ms RECORD
MAX MIN AVG
H
AC DC
DUTY
_ +
mVA
n F S %
k
%
_ +
0
1
2
3
4
5
6
7
8
9
0
40
mV
MIN MAX
RANGE
HOLD
H
REL
Hz
PEAK MIN MAX
mV
mA
V
A
V
A
OFF
mA A
A
COM
V
400mA MAX
FUSED
10A MAX
1000V MAX
FUSED
!
os19f.eps
4
4-19

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

863

Table of Contents