Ohms And Conductance Tests; Ohms Performance Test - Fluke 867B Service Manual

Hide thumbs Also See for 867B:
Table of Contents

Advertisement

867B,863
Service Manual

Ohms and Conductance Tests

Rotate the GMM selector to Ohms (
Configuration 5 (Figure 4-4). Then perform the test steps in Table 4-7 (for ohms) and
Table 4-8 (for conductance).
Before proceeding to the next test, place the 5700A in Standby.
Step
1
2
3
4
5
6
7
Step
1
2
4-10
Table 4-7. Ohms Performance Test
Range
Input
300.00Ω
0.00Ω
300.00Ω
190.00Ω
3.0000 kΩ
1.9000 kΩ
30.000 kΩ
19.000 kΩ
300.00 kΩ
190.00 kΩ
3.0000 MΩ
1.9000 MΩ
30.000 MΩ
10.000 MΩ
Table 4-8. Conductance Performance Test
Range
Input
300.00 nS
10.00 MΩ
3000.0 nS
1.0000 MΩ
ò
). Connect the GMM and the 5700A in
Input Source
Shorting Bar
5700A
5700A
5700A
5700A
5700A
5700A
Input Source
Lower Limit
5700A
99.30
5700A
993.0
Lower Limit
Upper Limit
0.00
0.10
189.77
190.23
1.8985
1.9015
18.985
19.015
189.85
190.15
1.8970
1.9031
9.977
10.023
Upper Limit
100.70
1007.0
4-9.