Dc Volts Performance Test; Dc Volts Test - Fluke 867B Service Manual

Hide thumbs Also See for 867B:
Table of Contents

Advertisement

867B,863
Service Manual

DC Volts Test

Connect the GMM and 5700A in Configuration 2 (Figure 4-2). Then perform the three
steps called for in Table 4-3. Before proceeding to the next test, place the 5700A in
Standby.
Step
1
2
3
1
DISPLAY
MODE
WAKE UP
4-6
Table 4-3. DC Volts Performance Test
GMM
Input
Range
30.000V
29.000V
300.00V
-290.00V
1000.0V
1000.0V
2
3
4
5
FREEZE
HZ
SAVE
MIN
RANGE
PRINT
MAX
TOUCH
HOLD
V
A
mA/ A
EXT TRIG
COM
V
Figure 4-2. Configuration 2 (DC Volts)
Model 867B
Lower
Input Source
Limit
5700A
28.991
5700A
-290.09
5700A
999.5
Model 863
Upper
Lower
Limit
Limit
29.009
28.986
-289.91
-290.14
1000.4
999.4
5700A Calibrator
OUTPUT
SENSE
V A
V
HI
LO
HI
AUX
GUARD GROUND
CURRENT
4-6.
Upper
Limit
29.014
-289.86
1000.6
WIDEBAND
HI
LO
os13f.eps

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

863