Performance Test; Testing The 1577 And 1587 Dmm Functions - Fluke 1587 Calibration Manual

Insulation multimeters
Hide thumbs Also See for 1587:
Table of Contents

Advertisement

1587/1577
Calibration

Performance Test

The following series of tests comprise a performance test for verifying the accuracy
of the Meter (UUT) and its performance level. The performance test is recommended
as an acceptance test for incoming inspection and as a calibration procedure for
periodically ensuring the accuracy of the Meter. Fluke recommends running the
performance test at least once a year.
No adjustments are required during the performance test, and it is not necessary to
open the case. If the Meter does not pass all parts of the performance tests, repair
and/or calibration adjustment are required. A calibration adjustment procedure is
given later in this manual. If significant repairs are required, contact Fluke as
described toward the front of this manual. If user repairs are appropriate, refer to the
list of user-replaceable parts toward the end of this manual

Testing the 1577 and 1587 DMM Functions

Use the following procedure to verify the accuracy of the measurement functions.
1. Connect the Calibrator to the J and COM input terminals on the UUT.
2. Turn the rotary switch to the function listed in each step of Table 2.
3. Apply the input level for each step listed in Table 2.
4. Compare the reading on the UUT display with the Display Reading in Table 2.
If the display reading falls outside of the range shown in Table 2, the UUT does
not meet specification.
UUT
Step
Function
1
mV ac
2
V ac
3
V ac
Push G (blue) for 1 second for low-pass filter band-pass test
4
V ac
5
V ac
Push G (blue) for 1 second to remove low-pass filter
6
V ac
7
V ac
8
V ac
9
V ac
10
Hz, V ac
11
Hz, V ac
12
Hz, V ac
13
V dc
14
V dc
15
V dc
16
Table 3. DMM Accuracy Tests
UUT
UUT
Range
Input Level
600.0 mV
30 mV, 60Hz
6.000V
3V, 60Hz
6.000V
3V, 400Hz
6.000V
3V, 400Hz
6.000V
3V, 800Hz
60.00V
60V, 50Hz
600.0V
600V, 50Hz
1000V
1000V, 50 Hz
1000V
1000V, 1 kHz,
99.99 Hz
*10 Hz, 0.15V
99.99 kHz
*95kHz, 0.15V
99.99 kHz
*90 kHz, 2V
6.000V
0.3V
6.000V
6V
60.00V
60V
UUT Display Reading
Fluke 1577
Fluke 1587
29.4 to 30.6
29.4 to 30.6
2.937 to 3.063
2.967 to 3.033
n/a
2.937 to 3.063
n/a
2.817 to 3.063
n/a
1.680 to 2.670
58.77 to 61.23
59.37 to 60.63
587.7 to 612.3
593.7 to 606.3
977 to 1023
n/a
977 to 1023
n/a
9.98 to 10.02
n/a
94.89 to 95.11
n/a
89.90 to 90.10
0.298 to 0.302
0.298 to 0.302
5.986 to 6.014
5.993 to 6.007
59.86 to 60.14
59.93 to 60.07
n/a