Accu-Chek Aviva Combo Owner's Booklet page 266

Hide thumbs Also See for Aviva Combo:
Table of Contents

Advertisement

Display Shows
Invalid PIN
Invalid Record
Time/Date
Invalid Target
Range
LO bG Warning
260
Possible Cause(s)
The wrong PIN was entered.
j
The time/date entered is
j
invalid (Add Data entries
cannot be set to occur in the
future).
The lower target range value
j
is above the upper target
range value.
The range selected conflicts
j
with your hyper and/or hypo
warning limit settings.
Your blood glucose may be
j
lower than the measuring
range of the system.
Possible Solution(s)
Select OK and re-enter the PIN shown on the pump
j
display.
Re-enter time/date.
j
Reset the target range values.
j
Enter the correct lower target range value and
j
upper target range value. See Chapter 5, Changing
Meter Settings.
Reset range or revise warning limit settings and
j
re-enter range.
Enter the correct lower target range value and
j
upper target range value. See Chapter 5, Changing
Meter Settings.
Treat your low blood glucose as recommended by
j
your healthcare professional.
j
If you are experiencing any of the common
symptoms of low blood glucose, contact your
healthcare professional immediately.