Display Shows
Bolus Delivery
Unavailable
Bolus Too High
Bolus Too Low
250
Possible Cause(s)
The meter cannot
j
communicate with the pump.
The pump is currently in Stop
j
mode.
The pump is currently
j
delivering a bolus.
The bolus is above the
j
allowed amount. The bolus
was set to 50 Units.
The insulin pump cannot
j
deliver a 0.1 Unit of multiwave
bolus.
Possible Solution(s)
Retry or set to Manual Pump.
j
Ensure the meter and pump have Bluetooth wireless
j
technology turned on.
Ensure the meter and pump are within
j
communication range.
j
If the low battery icon is displayed on the meter, the
meter cannot communicate with the pump. Replace
the batteries.
Take the pump out of Stop mode.
j
Allow the current bolus delivery to complete before
j
delivering the next bolus.
Check the accuracy of all entries. If necessary,
j
contact your healthcare professional.
The pump will adjust the bolus to 0.2 Units.