Description of Controls, Indicators, and Symbols
Pulse Search indicator. The Pulse Search indicator
displays before initial acquisition of a pulse signal and
throughout prolonged and challenged monitored
conditions. The Pulse Search indicator flashes
throughout a loss-of-pulse signal.
Data In-Sensor indicator. The Data In-Sensor
indicator displays to indicate that the attached O
sensor contains sensor-event data or the in-sensor
memory is full. The Data In-Sensor indicator flashes
when the sensor contains sensor-event data and when
sensor-event data is printed. The Data In-Sensor
indicator displays when the sensor memory is full.
Upper Alarm Limit indicator. The Upper Alarm Limit
indicator indicates that the displayed value is the upper
alarm limit for SpO
Lower Alarm Limit indicator. The Lower Alarm Limit
indicates that the displayed value is the lower alarm limit
for SpO
2
Data indicator. The Data indicator displays when the
NPB-40 is in the store snap-shot data or data print mode.
The Data indicator flashes when snap-shot data prints.
Print indicator. The Print indicator displays to indicate
that the NPB-40 is in the print mode. The Print indicator
flashes when the NPB-40 is prints single-event, snap-shot
or sensor-event data.