Performing Auto Calibration - Panasonic BT-LH910G Operating Instructions Manual

Lcd video monitor
Hide thumbs Also See for BT-LH910G:
Table of Contents

Advertisement

Main Menu
(continued)
• When using the CA-210 display color analyzer,
be sure to set the compensation values shown
on the right. If they are not set, correct calibration
will not be performed. For details on setup
procedures, refer to the CA-210 User's Guide.
There is no compensation when using Display Color Analyzer CA-310.
Turn on this unit and perform adequate aging (about 1 hour) before starting calibration.
Make the room dark so that no external light can enter the standard measurement probe before starting
the calibration. If external light enters the probe, the low brightness characteristics may not be calibrated
correctly.
LCD panel characteristics and instrument error in the display color analyzer may sometimes result in small
differences in values after calibration.
In a fine tuning of the monitor, also set GAIN and BIAS for R, G and B in the COLOR TEMP. VAR mode.
Do not apply the probe to a WFM/VECTOR display on the screen.
CALIBRATION
Select CALIBRATION in the "SYSTEM CONFIG" menu to open the following menus.
Sub menu
AUTO CALIBRATION
*
RESET
2
*
1
"EXECUTING" is displayed during "AUTO CALIBRATION" and "COMPLETE" appears when calibration
ends. "INCOMPLETE" appears if calibration could not be completed.
*
2
When "RESET" ends, "COMPLETE" appears.

Performing AUTO CALIBRATION

1.
Press [
][ ] in the "SYSTEM CONFIG"
menu, select the [CALIBRATION] submenu
and press [ENTER].
2.
Press [
][ ], select the [AUTO
CALIBRATION] submenu and press [ENTER].
[AUTO CALIBRATION]
MENU EXIT
36
Settings
Description
*
Connect a CA-210 Display color analyzer and use this
1
submenu to make a calibration.
Select "AUTO CALIBRATION" and select "YES" in the
confirmation screen that appears to start calibration.
Returns calibration data to their factory defaults.
Select "RESET" and select "YES" in the confirmation screen
that appears to return calibrated values to their factory defaults.
YES
NO
ENTER ENTER
SEL.
Compensation
W
x
0.296
y
0.3036
Lv
203.4
3.
Select "YES" and press [ENTER].
This starts calibration.
[AUTO CALIBRATION]
MENU EXIT
xxxxxxxxx indicates that one of the following
messages with the meaning listed below
appears.
EXECUTING: Operation in progress
COMPLETE: Operation completed
INCOMPLETE: Operation incompleted.
(Check the SERIAL connection or display color
analyzer connection.)
4.
Turn the power off and then on after
performing auto calibration.
R
G
0.6534
0.2851
0.1485
0.3271
0.5927
0.0523
43.18
148.1
xxxxxxxx
SEL.
ENTER ENTER
B
15.5