11.1.3 Testing the individual voltage of the batteries
Test the individual voltage U on the batteries that have been taken out by means of a Multimeter (measuring
range 50 V / 25 V DC (see chapter 10.1 page: 26).
Test
• Voltage under
12V?
• Voltage under
9.5V?
• Voltage under 0V? Þ Check fuses.
• Different voltage
(tolerance ± 0.2 V)
at each battery?
11.2

Check drive motor

11.2.1 Motor coil continuity

Remove drive unit panelling (see chapter 9.3 page: 19).
Remove the drive unit connecting plug at the controller (3).
Set the Multimeter to continuity check or resistance measuring.
Measure the continuity or resistance (R
between the red and black cable on the motor plug.
Test results:
No continuity in motor
circuit
(R = ∞ Ω).
Electra
Fault
Þ Battery not
completely charged.
Þ Battery fully
discharged.
The fuse is located
in the battery cable
bridges. (Chapter
10.1.5 page: 27)
Þ batteries defective
req
Broken cable in motor.
Carbon brushes worn
out.
Remedy
Þ Charge the battery with
®
an Invacare
charger.
Þ Pre-charge battery.
See chapter 12.1
page:35.
Þ "Ah Batter"
Check master fuse (50
A) of battery and cable /
connections, replace if
necessary, replace
battery if necessary
Þ Replace batteries
= 0 Ω) of the motor circuit
Þ Replace drive unit.
Þ Check / replace
carbon brushes.
Page 32
Service Manual
Check individual voltage
Drive unit connector
plug
Chapter:
Chapter: 9.6.5 page: 24
Chapter: 9.6.4 page: 24
As at:15.04.02