Error Codes - A&D GX-300 Instruction Manual

Gx series multi-function balance
Table of Contents

Advertisement

Weighing method
Does the weighing pan touch the breeze ring or anything? Is the weighing pan installed
correctly?
Is the RE-ZERO key pressed before placing a sample on the weighing pan?
Is the sample placed in the center of the weighing pan?
Has the balance been calibrated using the internal mass (one-touch calibration)?
Has the balance been warmed up for 30 minutes before weighing?
Sample and container
Has the sample absorbed or lost moisture due to the ambient conditions such as temperature
and humidity?
Has the temperature of the container been allowed to equalize to the ambient temperature?
Refer to "3-2 During Use".
Is the sample charged with static electricity? Refer to "3-2 During Use".
The GX-200/300/400/600/800/1000 is prone to be charged with static electricity when the
relative humidity is low.
Is the sample of magnetic material such as iron? Caution is required for weighing magnetic
materials. Refer to "3-2 During Use".

20-2 Error Codes

Display
Error code
Description
EC, E11
Stability error
The balance can not stabilize due to an environmental
problem.
changes, static electricity and magnetic fields.
Refer to "3. PRECAUTIONS" for details on the
operating
ADJUSTMENT" about adapting the balance to the
environment.
To return to the weighing mode, press the CAL key.
Out of range error
The value entered is beyond the settable range.
Re-enter the value.
EC, E16
Internal mass error
Applying the internal mass does not yield a change in
the weighing value as specified.
Confirm that there is nothing on the pan and perform
the weighing operation from the beginning again.
EC, E17
Internal mass error
The internal mass application mechanism does not
function properly.
Perform the weighing operation from the beginning
again.
Prevent
vibration,
environment
78
drafts,
temperature
and
"7.
RESPONSE

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents