Notes
If improper performance is found in the self check, the
balance displays
dealer for repair.
If the automatic response adjustment fails, the
balance displays
conditions such as breeze and vibration, also check
the weigning pan. Then, perform the adjustment again.
To return to the weighing mode, press the CAL key.
If the automatic response adjustment is awkward, try
to refine it using the manual response adjustment.

7-2 Manual Response Adjustment

This function manually updates the response adjustment.
Operation
1 Press and hold the MODE key until RESPONSE is
displayed, then release the key.
Press the MODE key again quickly.
2 Press the MODE key to select a rate of the response
adjustment. Either FAST, MID. or SLOW can be
selected.
3 After a few seconds of inactivity the balance displays
eend
. Then, it returns to the weighing mode and
displays
the
response indicator remains displayed for about 30
seconds.
Note
The response adjustment can be changed at "Condition
(Cond)" of "Environment, Display (ba5fnc)" in the
function table. For details, refer to "10. FUNCTION
TABLE".
CH no . Contact the local A&D
CH ng . Check the ambient
updated
response
indicator.
The
23
Press and hold
the key
Release the key
and press again
Each pressing
switches
the indicators
After a while