Trigger Subsystem; Abort; Initiate:continuous[:All]; Initiate[:Immediate][:All] - Agilent Technologies N5171B Command Reference Manual

X-series signal generators
Hide thumbs Also See for N5171B:
Table of Contents

Advertisement

System Commands

Trigger Subsystem

Trigger Subsystem

:ABORt

Supported
All
:ABORt
This command causes the List or Step sweep in progress to abort. If INIT:CONT[:ALL] is set to on,
the sweep will immediately re–initiate. The pending operation flag affecting *OPC, *OPC?, and *WAI
will undergo a transition once the sweep has been reset.

:INITiate:CONTinuous[:ALL]

Supported
All
:INITiate:CONTinuous[:ALL] ON|OFF|1|0
:INITiate:CONTinuous[:ALL]?
This command selects either a continuous or single list or step sweep. Execution of this command
does not affect a sweep in progress.
ON (1)
This choice selects continuous sweep where, after the completion of the previous
sweep, the current sweep will restart automatically or wait until the appropriate
trigger source is received.
OFF (0)
This choice selects a single sweep. Refer to
sweep triggering information.
*RST
0
Sweep Repeat Single Cont
Key Entry
Remarks
Execution of this command will not affect a sweep in progress.

:INITiate[:IMMediate][:ALL]

Supported
All
:INITiate[:IMMediate][:ALL]
This command either sets or sets and starts a single List or Step sweep, depending on the trigger
type. The command performs the following:
• arms a single sweep when BUS, EXTernal, or KEY is the trigger source selection
• arms and starts a single sweep when IMMediate is the trigger source selection
This command is ignored if a sweep is in progress. See
setting continuous or single sweep. See
source.
Single Sweep
Key Entry
188
:INITiate:CONTinuous[:ALL]
:TRIGger[:SEQuence]:SOURce
Agilent X-Series Signal Generators SCPI Command Reference
:INITiate[:IMMediate][:ALL]
command to select the trigger
for single
command for

Hide quick links:

Advertisement

Table of Contents
loading