Flatness:pair; Flatness:points; Flatness:preset; Flatness:step:points - Agilent Technologies N5171B Command Reference Manual

X-series signal generators
Hide thumbs Also See for N5171B:
Table of Contents

Advertisement

:FLATness:PAIR

Supported
All Models
[:SOURce]:CORRection:FLATness:PAIR <freq.>[<freq suffix>],
<corr.>[<corr suffix>]
This command sets a frequency and amplitude correction pair.
The maximum number of points that can be entered is 3,201.
<corr.>
This variable is the power correction.
Range
Frequency range varies and is model dependent. Refer to the instrument's Data
Sheet.
Key Entry
Configure Cal Array

:FLATness:POINts

Supported
All Models
[:SOURce]:CORRection:FLATness:POINts?
This query returns the number of points in the user–flatness correction file.

:FLATness:PRESet

Supported
All Models
CAUTION
The current correction data will be overwritten once this command is executed. Save the
current data if needed. Refer to
files.
[:SOURce]:CORRection:FLATness:PRESet
This command presets the user–flatness correction to a factory–defined setting that consists of one
point.
Preset List
Key Entry

:FLATness:STEP:POINts

Supported
All Models
[:SOURce]:CORRection:FLATness:STEP:POINts <points> |MAXimum|MINimum|DEFault|
[:SOURce]:CORRection:FLATness:STEP:POINts?[MAXimum|MINimum]
This command is used to define the number of points in the user flatness calibration step array.
See also
:FLATness:STEP:STARt
*RST
2
Range
2 to 3,201
# Points
Key Entry
Agilent X-Series Signal Generators SCPI Command Reference
:FLATness:STORe
and
:FLATness:STEP:STOP
Correction Subsystem ([:SOURce]:CORRection)
command for storing user–flatness
commands.
Basic Function Commands
21

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents