:SCLock:RATE

Supported
N5172B/82B
[:SOURce]:RADio:MTONe:ARB:SCLock:RATE <val>
[:SOURce]:RADio:MTONe:ARB:SCLock:RATE?
This command sets the sample clock rate for the multitone modulation format.
The variable <val> is expressed in units of hertz.
*RST
+60.000000E+006 (with Option 653)
+120.000000E+006 (with Option 655)
+80.000000E+006 (with Option 656)
+160.000000E+006 (with Option 657)
Range
Option 653: 1 Hz to 60 MHz
Option 655: 1 Hz to 120 MHz
Option 656: 1 Hz to 80 MHz
Option 657: 1 Hz to 160 MHz
ARB Sample Clock
Key Entry
Remarks
The modulation format should be active before executing this command. If this
command is executed before the modulation format is active, the entered value
will be overridden by a calculated factory default value. To activate the
modulation format, refer to

:SETup

Supported
N5172B/82B
[:SOURce]:RADio:MTONe:ARB:SETup "<file name>"
[:SOURce]:RADio:MTONe:ARB:SETup?
This command retrieves a multitone waveform file.
Key Entry
Load From Selected File
Remarks
The name of a multitone waveform file is stored in the signal generator file
system of MTONE files. This information is held in memory until you send the
command that turns the waveform on.
For information on the file name syntax, refer to
page

:SETup:STORe

Supported
N5172B/82B
[:SOURce]:RADio:MTONe:ARB:SETup:STORe "<file name>"
Agilent X-Series Signal Generators SCPI Command Reference
13.
Multitone Subsystem–N5172B/82B ([:SOURce]:RADio:MTONe:ARB)
"[:STATe]" on page
327.
Arb Commands
"File Name Variables" on
323

Hide quick links: