Programming the
Trigger Lines and
the Trigger Ports
Note
Enabling Trigger Lines
and the Trig Out Port
102
Triggering and System Status
The programming sequence used to set up the trigger lines and trigger ports
is shown in Figure 4-2. Detailed information on the commands used can be
found in Chapter 5 of this manual.
In the following commands, <n> is 0 or 1 when selecting a ECLTRG trigger
line, and 0 to 7 when selecting a TTLTRG* trigger line. Commands in
square brackets (
) are implied commands and are, therefore, optional.
[ ]
The brackets are not part of the command and are not sent to the instrument.
Figure 4-2. Backplane Trigger and Trig Out Port
In order to use a trigger line or the Trig Out port, the trigger line or port
must be enabled. This is done with the commands:
n
OUTPut:ECLTrg<
>[:STATe] 1 | ON
n
OUTPut:TTLTrg<
>[:STATe] 1 | ON
OUTPut:EXTernal[:STATe] 1 | ON
The reset condition for each of these commands is
trigger line or the Trig Out port must be enabled before it can be used.
Configuration Sequence
Enables a ECL trigger line
Enables a TTL trigger line
Enables the Trig Out port
. Therefore, a
OFF
.
.
.
Chapter 4