Analog Modulation Commands
Phase Modulation Subsystem–Option UNT ([:SOURce])
Remarks
The RF carrier is modulated when you set the signal generator's modulation state
to ON, see
Whenever phase modulation is enabled, the M annunciator is turned on in the
display

:PM[1]|2[:DEViation]

Supported
All Models with Option UNT
[:SOURce]:PM[1]|2[:DEViation] <value><unit>|UP|DOWN
[:SOURce]:PM[1]|2[:DEViation]?
This command sets the deviation of the phase modulation.
The variable <unit> will accept RAD (radians), PIRAD (pi- radians), and DEG (degrees); however, the
query will only return values in radians.
0.00000000E000
*RST
Range
See the Data Sheet for range values
M Dev
Key Entry
Remarks
Refer to
associated with the UP and DOWN choices.

:PM[:DEViation]:STEP[:INCRement]

Supported
All Models with Option UNT
[:SOURce]:PM[:DEViation]:STEP[:INCRement] <value><unit>
[:SOURce]:PM[:DEViation]:STEP[:INCRement]?
This command sets the phase modulation deviation step increment.
Range
0.001–1E3RAD
Incr Set
Key Entry
Remarks
The value set by this command is used with the UP and DOWN choices for the FM
deviation command. Refer to
The setting enabled by this command is not affected by signal generator power- on,
preset, or *RST.
236
:MODulation[:STATe]
:PM[:DEViation]:STEP[:INCRement]
:PM[1]|2[:DEViation]
command for more information.
command for setting the value
command for more information.