Arb Commands
Dual ARB Subsystem–N5172B/82B ([:SOURce]:RADio:ARB)
would make the first marker point occur after the last, the last marker point is
automatically adjusted to match the first marker point.
<last_point>
This variable defines the last point in the range over which the marker will be
placed. This value must be greater than or equal to the first point, and less than
or equal to the total number of waveform points.
<skip_count>
This variable defines the marker point pattern across the range. A zero value
means the marker points occur consecutively across the range. A value greater
than zero creates a repeating marker point pattern across the range, where the
gap between the marker points is equal to the <skip_count> value. The gaps begin
after the first marker point. Each marker point in the pattern, which is only one
point wide, produces a marker pulse.
Example
:RAD:ARB:MARK "Test_Data",1,40,100,2
The preceding example sets marker 1 on the first point, 40, the last point, 100, and every third point
(skip 2) between 40 and 100 (assuming the Test_Data file has at least 100 points).
Range
<marker>: 1–4
<first_Point>: 1–number of waveform points
<last_point>: <first_Point>–number of waveform points
<skip_count>: 0–number of points in the range
Set Marker on Range Of Points
Key Entry
# Skipped Points

:MBSync

Supported
N5172B/82B
[:SOURce]:RADio:ARB:MBSync OFF|MASTer|SLAVe
[:SOURce]:RADio:ARB:MBSync?
This command disables or enables the multiple BBG synchronization setup for the current signal
generator. Ensure that the Dual ARB play is off prior to executing this command.
OFF
Turns off multiple baseband synchronization for the signal generator.
MASTer
Sets the signal generator as the master for the setup. When selected, the following
trigger features are unavailable:
Trigger Type
Prior to selecting MASTer, ensure that the trigger type is something other than
shown above. If not, the X- Series signal generator generates a settings conflict
error and changes the trigger type to TRIGger (continuous play once triggered).
284
Apply to Waveform
Free Run, see
page 303
Gated, see
page 302
Marker 1 2 3 4
First Mkr Point
Agilent X-Series Signal Generators SCPI Command Reference
Last Mkr Point