Rtd Types Accepted - Fluke 726 User Manual

Multifunction process calibrator
Hide thumbs Also See for 726:
Table of Contents

Advertisement

RTD Type
Pt100 (3926)
Pt100 (385)
Ni120 (672)
Pt200 (385)
Pt500 (385)
Pt1000 (385)
Pt100 (3916)
The IEC standard RTD commonly used in U.S. industrial applications is the Pt100 (385),
Pt100 (3916),
= 0.003916
Custom RTDs may also be added, see PRT Custom Curves
Table 6. RTD Types Accepted
Ice Point (R
)
0
100
Platinum
100
Platinum
120
Nickel
200
Platinum
500
Platinum
1000
Platinum
100
Platinum
/ C is also designated as JIS curve.
Material
0.003926
0.00385
0.00672
0.00385
0.00385
0.00385
0.003916
Multifunction Process Calibrator
Using Measure Mode
Range ( C)
/ C
-200 to 630
/ C
-200 to 800
/ C
-80 to 260
/ C
-200 to 630
/ C
-200 to 630
/ C
-200 to 630
/ C
-200 to 630
= 0.00385
/ C.
25

Hide quick links: