Simulating Rtds - Fluke 726 User Manual

Multifunction process calibrator
Hide thumbs Also See for 726:
Table of Contents

Advertisement

726
Users Manual

Simulating RTDs

Connect the Calibrator to the instrument under test as
shown in Figure 17. Proceed as follows to simulate an
RTD:
M
1.
If necessary, press
R
2.
Press
for the RTD display.
Note
Use the 3W and 4W terminals for measurement
only, not for simulation. The Calibrator simulates
a 2-wire RTD at its front panel. To connect to a
3-wire or 4-wire transmitter, use the stacking
cables to provide the extra wires. See Figure 17.
36
for SOURCE mode.
3.
Enter the desired temperature by pressing
W
Y
Z
. Press
and
to select a different digit to edit.
4.
If the 726 display indicates ExI HI, the excitation
current from your device under test exceeds the
limits of the 726.