Specifications

Specifications are based on a one year calibration cycle and apply from +18 C to +28 C unless stated otherwise. All
specifications assume a five-minute warmup period.

DC Voltage Measurement and Source

Range
30 V (upper display)
20 V (lower display)
20 V (Source)
100 mV (Source)
90 mV (Read)
Maximum current output in voltage ranges is 1 mA with an output impedance of

DC mA Measurement and Source

Range
mA Read (Upper Display)
mA Read (Lower Display)
mA Source
Maximum load on, mA source is 1 ke. With the HART resistor on, maximum load is 750 .
Voltage input range on simulate mode is 5 to 30 V
Minimum
0.000
0.000
0.000
0.000
0.000
Minimum
0.000
0.000
0.000
Maximum
30.000
20.000
20.000
100.000
90.000
1
Maximum
24.000
24.000
24.000
Multifunction Process Calibrator
Specifications
Accuracy,
(% of Reading + Floor)
0.010 % + 2 mV
0.010 % + 2 mV
0.010 % + 2 mV
0.010 % + 10 µV
0.010 % + 10 µV
Accuracy,
(% of Reading + Floor)
0.010 % + 2 A
0.010 % + 2 A
0.010 % + 2 A
59