Set The Sweep Start Value To 100 Hz; Set The Sweep Stop Value To 100 Mhz; Set The Measurement Bandwidth To 2 - Agilent Technologies 4294A Operating Manual

Precision impedance analyzer
Hide thumbs Also See for 4294A:
Table of Contents

Advertisement

Learning Operation Basics
Specifying Measurement Conditions
Step 2. Press the
key to select Log (logarithmic) sweep.
LOG

Set the Sweep Start Value to 100 Hz

Step 1. Press the
key. The current setting of the sweep start value appears in the Parameter
[Start]
Setting field in the upper-left area of the screen.
Step 2. Type "100" into the Parameter Setting field using these ENTRY block keys:
Step 3. Specify that the value does not take any unit by pressing the
This puts your entry into effect.

Set the Sweep Stop Value to 100 MHz

Step 1. Press the
key. The current setting of the sweep stop value appears in the
[Stop]
Measurement Parameter field in the upper-left area of the screen.
Step 2. Type "100" into the Parameter Setting field using these ENTRY block keys:
Step 3. Suffix your entry with "M" (mega) by pressing the
puts your entry into effect.

Set the Measurement Bandwidth to 2

Step 1. Press the
[Bw/Avg]
Step 2. Press the
BANDWIDTH [ ]
Step 3. Set the measurement bandwidth to 2 by pressing the
34
key to display the Measurement Bandwidth/Averaging menu.
key to display the Measurement Bandwidth Setting menu.
[1][0][0]
key in the ENTRY block.
[´1]
[1][0][0]
key in the ENTRY block. This
[M/m]
key.
2
.
.
Chapter 2