Page:measure[:Sweep]:Var1:Start - Agilent Technologies 4155C Command Reference Manual

Semiconductor parameter analyzer precision semiconductor parameter analyzer
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

Syntax
Parameter
Query response
Example
Syntax
Parameter
SCPI Commands

:PAGE:MEASure[:SWEep]:VAR1:STARt

:PAGE:MEASure[:SWEep]:VAR1:STARt
This command sets the sweep START value of VAR1.
At *RST, this value is 0 V.
:PAGE:MEASure [ :SWEep] :VAR1:STARt start | MINimum | MAXimum
Parameter
Type
start
numeric
start <newline><^END>
start is NR3 response data type.
OUTPUT @Hp4155;":PAGE:MEAS:VAR1:START 0"
OUTPUT @Hp4155;":PAGE:MEAS:VAR1:START MIN"
OUTPUT @Hp4155;":PAGE:MEAS:VAR1:START?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:MEAS:VAR1:START? MIN"
ENTER @Hp4155;A
:PAGE:MEASure[:SWEep]:VAR1:STEP
This command sets the sweep STEP value of VAR1 for the linear sweep. This
parameter is not used for logarithmic sweep.
At *RST, this value is 0.01 V.
:PAGE:MEASure [ :SWEep] :VAR1:STEP step | MINimum | MAXimum
Parameter
Type
step
numeric
5-230
Agilent 4155C/4156C SCPI Command Reference, Edition 1
Explanation
-200 to 200 V or -1 to 1 A. The range of this value
depends on the unit type of VAR1.
Explanation
-400 to 400 V or -2 to 2 A. The range of this value
depends on the unit type of VAR1.