To set the salinity value (SALI Value)
Auto power off
3200D Dissolved Oxygen Meter Operating Guide
You do not usually need to set salinity. The default salinity value
is 0.0 g/L.
1 From the initial state, press [Setup]. The setup screen
appears. See
Figure 12
2 Use the directional arrow keys to navigate to SALI Value.
3 Press [Enter] to select SALI Value or press [Cancel] to exit.
The SALI Value screen appears on the display. Enter the salinity
pressure.
Figure 19
Set salinity value
The meter can power off automatically after a set time period.
To set the Auto Power Off time:
1 From the initial state, press [Setup]. The setup screen
appears. See
Figure 12
2 Use the directional arrow keys to navigate to Auto Power Off.
3 Press [Enter] to select Auto Power Off or press [Cancel] to exit.
4 The Auto Power Off screen appears on the display. Enter a
power down time from 10 to 480 minutes. Enter a time of 0 to
disable Auto Power Off.
5 Press [Enter].
When the Auto Power Off time elapses after the start of a
measurement, the meter shuts down. If the meter is connected
to the Electrochemical Data Collecting Software, Auto Power Off
is disabled.
on page 26.
on page 26.
Operation
2
35