STATus:QUEStionable:UMCounter:TREMaining?
Returned Data
Comment
Reset Condition
Related
Commands
Chapter 3
The
STATus:QUEStionable:UMCounter:TREMaining
representing the time remaining on the user maintenance counter. The user
maintenance counter is a countdown timer that counts hours down to zero. The time
remaining is the number of hours until a UMCounter warning is issued.
Type
Range
Default
int16
0 - 65535
65535
•
The value 65535 means that the counter is disabled. It is shipped from the
factory disabled. Set the time interval (
the counter (
STAT:QUES:UMC:TRES
*RST does not affect the status system.
STATus:QUEStionable:UMCounter:TINTerval,
STATus:QUEStionable:UMCounter:TREMaining?
Description
number of hours remaining in the UMCounter.
STAT:QUES:UMC:TINT
) to enable the user maintenance counter.
Programming the Enhanced Monitor
query returns an integer
) and then reset
117