f) Dampen a cotton swab or lint free cloth with the
rinse solvent and gently rub the optical surface.
g) Air-dry the probe for a few minutes.
h) Reinsert into the process and follow the required
initialization sequence (see above)
4) If Step 1-3 are not sufficient to return the Fluorometer to
operation, contact TIA for assistance.
5.8.3

Rectifying the Problem: Red Indicators

If there has been a major change in the probe installation or process
piping. Most likely, there is a stray light leak into the system. Verify
that the monitoring location is in complete darkness. If stray light is not
present at the monitoring location, contact TIA for assistance.