Operational Theory; Zro 2 Sensor; Principles Of Operation - Teledyne 3000ZA-3X Operating Instructions Manual

Trace oxygen analyzer
Table of Contents

Advertisement

Trace Oxygen Analyzer

Operational Theory

2.1 Introduction
The analyzer is composed of three subsystems:
1. Three Zirconium Oxide (ZrO
specific range of analysis: Low range sensor of 0-200
ppm, medium range sensor of 0-2 %, and high range
sensor of 0-95%.
2. Sample System
3. Electronic Signal Processing, Display and Control
The sample system is designed to accept the sample gas and
transport it through the analyzer without contaminating or altering the
sample prior to analysis. The ZrO
amount of oxygen present in the sample into an electrical current. The
electronic signal processing, display and control subsystem simplifies
operation of the analyzer and accurately processes the sampled data. The
microprocessor controls all signal processing, input/output and display
functions for the analyzer.
2.2 ZrO
Sensor
2

2.2.1 Principles of Operation

Teledyne's zirconium oxide sensor is a miniature solid state sensor.
It uses a stabilized zirconia disc as an electrolyte with a sensing
electrode (the cathode) and a counter electrode (the anode) on each side
of the disc. The zirconia is heated to a temperature of about 500°C via a
built-in heater. By applying a suitable reference voltage (approximately
0.75 volt) across the cathode and anode, the heated zirconia acts as an
electrolyte which is capable of moving oxygen ions within its crystalline
structure. This allows for the reduction of oxygen molecules reaching
the cathode and the generation of a current equivalent to the parts per
million oxygen in the sample gas.
Teledyne Analytical Instruments
Operational Theory
) Sensors. Each one has a
2
sensor is a device that translates the
2
7

Advertisement

Table of Contents
loading

Table of Contents