Span Sensor; Span Value; Span Timer - Teledyne 3000ZA-3X Operating Instructions Manual

Trace oxygen analyzer
Table of Contents

Advertisement

Trace Oxygen Analyzer

4.4.1 Span Sensor

-----------------------
->Span Sensor: High
-----------------------
Pressing ENTER causes the selection after 'Sensor:' to start flashing
(Low / Mid / High). Then pressing UP or DOWN selects a different
sensor. Note that the span value directly below will change according to
which sensor is currently selected.
The high sensor covers ranges from 2 to 95% O
near the top of this scale should be used to calibrate this sensor.
The medium sensor covers ranges from 200 ppm to 2% O
gas that is near the top of this scale should be used to calibrate this
sensor.
The low sensor covers ranges from 0 to 100 ppm O
is near the top of this scale should be used to calibrate this sensor.

4.4.2 Span Value

-----------------------
->Span 85%
-----------------------
Pressing ENTER causes the number after 'Span' to start flashing.
The number has a range of 1 ppm to 100% and should be set equal to the
concentration of the span gas used. This number corresponds to the
sensor selected above. Pressing UP or DOWN adjusts the number.

4.4.3 Span Timer

-----------------------
->Span Timer: 7 min.
-----------------------
This function sets the hold time in span mode before the analyzer
actually performs the span. Adjust to give enough time to purge with
span gas.
Teledyne Analytical Instruments
Operation
. Span gas that is
2
. Span
2
. Span gas that
2
37

Advertisement

Table of Contents
loading

Table of Contents