Cleaning; Troubleshooting; Table 5-2: Troubleshooting - Teledyne 3000ZA-3X Operating Instructions Manual

Trace oxygen analyzer
Table of Contents

Advertisement

Maintenance
Sample system
Power Supply
Microprocessor
Displays
5 digit LED meter
2 line, 20 character, alphanumeric, VFD display
RS-232 Communications Port
See the drawings in the Drawings section in back of this manual for
details.

5.6 Cleaning

If instrument is unmounted at time of cleaning, disconnect the
instrument from the power source. Close and latch the front-panel access
door. Clean outside surfaces with a soft cloth dampened slightly with plain
clean water. Do not use any harsh solvents such as paint thinner or benzene.
For panel-mounted instruments, clean the front panel as prescribed
in the above paragraph. DO NOT wipe front panel while the instrument
is controlling your process.

5.7 Troubleshooting

Table 5-2: Troubleshooting

Problem
Erratic readings of the
oxygen concentration
as reported by the
analyzer
Teledyne Analytical Instruments
Possible Cause
Analyzer may have
been calibrated in an
inaccurate fashion.
3000ZA-3X
Solution
Recalibrate analyzer,
making sure the
proper gas is fed, the
proper sensor is
selected, and the
proper value of span
gas is input.
50