Manual Range Select; Range Setpoints - Teledyne 3000ZA-3X Operating Instructions Manual

Trace oxygen analyzer
Table of Contents

Advertisement

Operation

4.7.2 Manual Range Select

-----------------------
Man Rng:
-----------------------
Pressing ENTER causes the range R1 or R2 or R3 to start flashing.
Then pressing UP or DOWN cycles through the 3 ranges. The selected
range is the effective range if the Range Select (above) is set to
Man(ual).

4.7.3 Range Setpoints

-----------------------
R1: 0-1000.00 ppm
-----------------------
Pressing ENTER causes the value to the right of Rn: to start
flashing. Then pressing UP or DOWN adjusts the value. These are the
range setpoints for R1, R2, and R3. They may be set to a value in the
range of 1.00 ppm to 100% with the stipulation that the each range must
have a setpoint equal to or greater than the setpoint of the range below it
plus 1ppm.
Teledyne Analytical Instruments
R1
3000ZA-3X
44

Advertisement

Table of Contents
loading

Table of Contents