Logout; System Self-Diagnostic Test - Teledyne 3000TA Operating Instructions Manual

Trace oxygen analyzer
Hide thumbs Also See for 3000TA:
Table of Contents

Advertisement

Trace Oxygen Analyzer
will immediately switch to the Analyze screen, and you now have access to
all instrument functions.
If all alarms are defeated, the Analyze screen appears as:
If an alarm is tripped, the second line will change to show which alarm
it is:
NOTE: If you log off the system using the logout function in the system
menu, you will now be required to re-enter the password to gain
access to Span, Zero, Alarm, and Range functions.

4.3.4 Logout

The Logout function provides a convenient means of leaving the
analyzer in a password protected mode without having to shut the instrument
off. By entering Logout, you effectively log off the instrument leaving the
system protected against use until the password is reentered. To log out,
press the System button to enter the Syste`m function.
Use the < > arrow keys to position the blinking over the Logout func-
tion, and press Enter to Log out. The screen will display the message:

4.3.5 System Self-Diagnostic Test

The Model 3000TA has a built-in self-diagnostic testing routine. Pre-
programmed signals are sent through the power supply, output board and
sensor circuit. The return signal is analyzed, and at the end of the test the
status of each function is displayed on the screen, either as OK or as a
number between 1 and 3. (See System Self Diagnostic Test in chapter 5 for
number code.)
The self diagnostics are run automatically by the analyzer whenever the
instrument is turned on, but the test can also be run by the operator at will.
To initiate a self diagnostic test during operation:
0.0
ppm Anlz
Range: 0 — 100
0.0
ppm Anlz
AL—1
TRAK/HLD Auto—Cal
PSWD Logout More
Protected Until
Password Reentered
Teledyne Analytical Instruments
Operation 4
4-9

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents