Trace Oxygen Analyzer
1. Remove power to the instrument by unplugging the power cord
at the power source.
2. Open the front panel door by pressing the release button on the
top right corner of the door all the way in with a narrow gauge
tool, such as a small screwdriver, and releasing it.
3. With one hand placed underneath the cell block ready to catch
the Micro-Fuel cell , lift up on the stainless steel gate in front of
the cell block. This releases the cell and cell holder from the
block. The cell and holder will fall out in your hand.
Lift Up
Gate
Figure 5-1: Removing the Micro-Fuel
Teledyne Analytical Instruments
Maintenance 5
Cell Block
Micro-Fuel Cell
Cell Adaptor
(For B-2 or A-2
series cell)
O-Ring
Cell Holder
5-3