Trace Oxygen Analyzer
CALIBRATION GAS:
3000 series analyzer requirements for units with Auto-Cal options. The
customer must supply a control valves (or restrictors) for any SPAN or ZERO gas
source which is attached to the Auto-Cal ports. The valve should be adjusted to the
same flow rate as the sample gas . When restrictors are used, the gas pressure must
be adjusted to achieve the proper flow rate.
OPERATION WITHOUT A RESTRICTOR DEVICE:
Operation without a restrictor device is not recommend as mentioned
above. A 3000TA without any flow restrictor device was tested on 11-19-97.
This results in a flow rate of 2.4 SLPM @ 1 PSIG. This is a cv of 0.023 for the
standard sample sys.
REFERENCE: FLOW_1.XLS & FLOW_2.XLS for information on flow rates
at various pressures.
TAI PART NUMBERS
RESTRICTOR KIT:
UNION (SS)
LP. RESTRICTOR
STD.. RESTRICTOR
NUT
FERRULE
FERRULE
CONVERSIONS:
1 PSI
1 SCFH
A68729
U11
R2323 ( LOW PRESSURE / VAC. SERVICE )
R2324
N73
F73
F74
=
2.04 INCHES OF MERCURY (in. Hg.)
=
0.476 SLPM
Teledyne Analytical Instruments
BLUE DOT
BOTH FERRULES ARE
REQUIRED
Appendix
A-7