Teledyne 3000TA Operating Instructions Manual page 50

Trace oxygen analyzer
Hide thumbs Also See for 3000TA:
Table of Contents

Advertisement

4 Operation
If you are using password protection, you will need to enter your
password to access the alarm functions. Follow the instructions in section
4.3.3 to enter your password. Once you have clearance to proceed, enter the
Alarm function.
Press the Alarm button on the front panel to enter the Alarm function.
Make sure that AL–1 is blinking.
Set up alarm 1 by moving the blinking over to AL–1 using the < >
arrow keys. Then press Enter to move to the next screen.
Five parameters can be changed on this screen:
Value of the alarm setpoint, AL–1 #### ppm (oxygen)
Out-of-range direction, HI or LO
Defeated? Dft–Y/N (Yes/No)
Failsafe? Fs–Y/N (Yes/No)
Latching? Ltch–Y/N (Yes/No).
To define the setpoint, use the < > arrow keys to move the
blinking over to AL–1 ####. Then use the Δ∇ arrow keys to
change the number. Holding down the key speeds up the
incrementing or decrementing. (Remember, the setpoint units are
ppm O
2
To set the other parameters use the < > arrow keys to move the
blinking over to the desired parameter. Then use the Δ∇ arrow
keys to change the parameter.
Once the parameters for alarm 1 have been set, press Alarms
again, and repeat this procedure for alarm 2 (AL–2).
To reset a latched alarm, go to Dft– and then press either Δ two
times or ∇ two times. (Toggle it to Y and then back to N.)
Go to Ltch– and then press either Δ two times or ∇ two times.
(Toggle it to N and back to Y.)
4-18
AL—1
Choose Alarm
AL—1 1000 ppm HI
Dft—N Fs—N Ltch—N
.)
–OR –
Teledyne Analytical Instruments
AL—2
Model 3000TA

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents