Agent class
Water
Cleaning agent
Alkaline solution
Note
Information on cleaning or disinfection of other system components can be obtained from
the respective instructions for use.

8.1.2 Prohibited cleaning agents and disinfectants

The following cleaning agents and disinfectants can bleach the paint after a longer period of
application:
Agent class
Light gasoline

8.2 Check the settings

CAUTION
Checking the settings
• The settings may only be checked by service personnel.
• The settings must not be checked in the presence of patients.
The picture quality of the monitor changes due to aging of the LCD unit and the backlight.
Check the monitor settings at regular intervals in accordance with the local guidelines.
Correct the settings if necessary.
CuratOR LX1910
Instructions for Use, 08/2022
Cleaning and check settings
Tested cleaning agents and disinfectants
Tap water
Distilled water
Ammonium solution 1,65 vol. %
Limewash, saturated Ca(OH)
Tested cleaning agents and disinfectants
Petroleum spirit
Petroleum ether
8.2 Check the settings
-solution