Function
LUT
Signal
Timing Change Delay
(ms)
Image Reset
CuratOR LX1910
Instructions for Use, 08/2022
Values
1 - DICOM (0.6–400)
2 - DICOM (0.8–400)
3 - CRT (0.8–400)
4 - CIE DIN6174 (0.8–400)
5 - Gamma 2.2 (0.8–400)
No Correction
Default: 1 - DICOM (0.6-400)
Color RGB
Monochrome (RGB)
Monochrome (R)
Monochrome (G)
Monochrome (B)
Default: Color RGB
Inactive
50–500
Default: Inactive
Description
Select the Look Up Table (LUT)
The LUT determines the monitor's gamma
curve. By using a different LUT for exam-
ple, you can highlight specific grayscale
levels.
Note: Select a DICOM LUT to view radio-
graphic images.
Note: If "No Corretionc" is selected, the
functions "Color Temperature" and "Color
Fine Adjustment" are not available.
Switch signal between color and mono-
chrome
Color RGB: Color setting
Monochrome (RGB): Red, green and
blue are displayed as grayscale
weights.
Monochrome (R): The red channel is
displayed as a monochrome signal.
Monochrome (G): The green channel is
displayed as a monochrome signal.
Monochrome (B): The blue channel is
displayed as a monochrome signal.
Set the delay for timing changes
Changing one of the following parameters
corresponds to a timing change, and the
monitor begins to resynchronize ("Auto in
progress" displayed):
H frequency
V frequency
V total
To prevent this from happening as a result
of each and every signal disturbance, the
value representing the permissible number
of faulty or changed borders can be in-
creased using the "Signal change delay"
function. This is required in the case of un-
stable signal sources, for example.
Note: Please note, however, that once the
function is active, there will be a delay of
several milliseconds even for desired tim-
ing changes. The higher the value of the
function, the greater the delay.
Reset the image parameter to the de-
fault settings
All settings made in the "Image" menu, ex-
cept for "color temperature", are reset to
the defaults. The monitor settings and set-
tings of the AD converter remain in place.