Note
Information on cleaning or disinfection of other system components can be obtained from
the respective instructions for use.

8.1.2 Prohibited cleaning agents and disinfectants

The following cleaning agents and disinfectants can bleach the paint after a longer period of
application:
Agent class
Peroxide compounds
Petroleum spirit

8.2 Check settings

CAUTION
Checking the settings
• The settings may only be checked by service personnel.
• The settings must not be checked in the presence of patients.
Check the settings on a regular basis
The picture quality of the monitor changes due to aging of the LCD unit and the backlight.
Check the monitor settings at regular intervals in accordance with the local guidelines.
Correct the settings if necessary.
CuratOR EX190
Instructions for Use, 06/2022
Tested cleaning agents and disinfectants
Perform
Petroleum spirit close to boiling
Cleaning and check settings
8.2 Check settings
Further examples
Dismozon pur
Petroleum ether