Appendix
12.2 Information on electromagnetic compatibility (EMC)
Electromagnetic interference immunity
The CuratOR LX1910 was tested with the following compliance levels in accordance with the test re-
quirements for professional healthcare facilities, as established in IEC/EN 6061-1-2.
Customers and users of the CuratOR LX1910 have to ensure that the monitor is used in such an en-
vironment.
Interference im-
Measurement
munity test
level
Electrostatic dis-
±8 kV contact
charge (ESD)
±15 kV air
IEC/EN
61000-4-2
Fast transient
±2 kV power lines
electric distur-
±1 kV input / out-
bances (bursts)
put lines
IEC/EN
61000-4-4
Surge voltage
±1 kV line
against line
IEC/EN
61000-4-5
±2 kV line against
ground
Voltage dips,
0 % V
brief interrup-
riods and 1 period
tions, and fluctu-
70 % V
ations of power
30 periods at 50 /
supply lines
60 Hz
IEC/EN
0 % V
61000-4-11
300 periods at
50 / 60Hz
Magnetic fields
30 A/m
with energy
(50 / 60 Hz)
technology fre-
quencies
IEC/EN
61000-4-8
Note: V
is the alternating current voltage before application of the measurement level.
T
46
Compliance level
±8 kV contact
±15 kV air
±3 kV power lines
±2 kV input / out-
put lines
±2 kV line against
line
±4 kV line against
ground
for 0.5 pe-
0 % V
for 0.5 peri-
T
T
ods and 1 period
for 25 /
70 % V
for 25 pe-
T
T
riods at 50 Hz
0 % V
for 250 pe-
T
for 250 /
riods at 50 Hz
T
30 A/m
(50 Hz)
Information regarding the electromag-
netic environment
It is recommended to use the device on
wood, concrete, or ceramic floors. If the
floor is made of synthetic material, the
relative humidity should be at least 30%.
The power supply quality has to corre-
spond to that of typical industrial environ-
ments or hospitals.
The power supply quality has to corre-
spond to that of typical industrial environ-
ments or hospitals.
The power supply quality has to corre-
spond to that of typical industrial environ-
ments or hospitals.
If the monitor has to continue operation
even if the power supply is interrupted, it
is recommended to connect the device to
an uninterruptible power supply or bat-
tery.
The magnetic fields with energy technol-
ogy frequencies must be in an area that
is representative of a typical location in a
typical industrial environment or hospi-
tals.
The product should be kept at least
15 cm away from the source of power
frequency magnetic fields during use.
Instructions for Use, 08/2022